Procedure - Solid Samples
Instrument calibration/drift correction.
Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 3b through 3j a minimum of three times for each calibration/drift standard used.
Calibrate/drift following the procedure outlined in the operator's instruction manual.
Instrument calibration/drift correction
f. Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner.
g. Firmly place a graphite crucible on the lower electrode tip.
h. Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
i. Repeat steps 3b through 3h a minimum of three times for each calibration/drift standard used.
. Calibrate/drift following the procedure outlined in the operator's instruction manual.
Method Selection
Two methods are described in this application note; either method can be used to analyze iron, steel, nickel-, and cobalt-base alloys. The Precision Method is recommended for general use and will provide the best precision and accuracy throughout the typical O, N, and H concentrations found in this group of metals; approximate cycle time is 3.5 minutes. The Fast Track method can be used where speed of analysis is a critical component; for example, when molten metal is being sampled and results are required in the shortest possible time. This method will produce suitable results for most samples; approximate cycle time is 2.25 minutes. As noted above, sampling and sample preparation are key elements to accurate O, N, and Hdetermination as well. It is up to the user to determine which method best meets their needs.
Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
Place the Nickel Capsule containing the sample into the open port at the top of the loading head.
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode either manually or remove the crucible and press the analyze button to clean with an automatic cleaner if applicable.
ONH836氧氮氢分析仪
高频燃烧红外检测技分析无机材料中的碳、硫含量。
采用惰性气体保护熔融法,提供各种无机材料、钢铁、有色金属、硬质合金、难熔金属、陶瓷等材料中宽范围的氧氮氢同时分析。
该仪器配置MS Windows®的软件接口,专为触摸屏操作定制。长期的开发研究与客户反馈相结合,这个简单方便的软件使得用户可以快速完整访问仪器控制,分析设置,故障诊断,数据报告, **提高日常分析的效率并节省用户宝贵的时间。 可选的Cornerstone® Mobile软件功能,使得用户可以在手机,便携式电脑及台式电脑上便捷的访问仪器数据,分析曲线图形以及仪器的状态,实现对仪器的遥控。 OHN836系列仪器采用惰性熔融法。湖南LECO氧氮氢分析仪采购
国际**的非色散红外及热导检测器 ,无移动部件,不需要手动调整。湖南LECO氧氮氢分析仪采购
Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Calibrate/drift following the procedure outlined in the operator's instruction manual.
Login Sample with the appropriate number of reps.
Weigh approximately *** grams of appropriately prepared sample into a *** Nickel Capsule, enter mass and sample identification into appropriate rep fields. If applicable, place the nickel capsule in the
appropriate autoloader position.
Repeat steps 3c through 3i for sample analysis.
上海禹重实业有限公司是一家从事仪器科技、医疗科技、检测科技领域内的技术咨询、技术服务,仪器仪表的修理,仪器仪表、实验室设备、机械设备及配件、机电设备及配件、电子产品、化工产品及原料(除危险化学品、监控化学品、烟花爆竹、民用易爆物品、易制毒化学品)、第二类医疗器械的销售,机械设备租赁,电子商务,货物进出口,技术进出口。的公司,是一家集研发、设计、生产和销售为一体的专业化公司。公司自创立以来,投身于光谱仪,色谱仪,质谱仪,样品前处理,是仪器仪表的主力军。禹重科技始终以本分踏实的精神和必胜的信念,影响并带动团队取得成功。禹重科技始终关注自身,在风云变化的时代,对自身的建设毫不懈怠,高度的专注与执着使禹重科技在行业的从容而自信。
氧氮氢分析仪的常见故障及解决方法仪器启动时显示。没有水流。系统分析电流切断,分析停止。这是水流探测器...
【详情】736Series:Oxygen/NitrogenbyInertGasFusionThe736fam
【详情】美国LECO力可氧氮氢分析仪ONH836氧氮氢联测仪系列软件部分1、用户友好界面的Cornersto...
【详情】Procedure–Powder/ChipSamplesInstrumentcalibration/...
【详情】LECO—YoursourcefortotalanalyticalsolutionsCSCS844a...
【详情】InstrumentHighlightsandFeaturesFeaturesandBenefits
【详情】仪器的亮点和优势:特点和优势·炉头区域LED照明系统·炉头快速拆装的粉尘过滤装置·装有自动旁路阀的可...
【详情】Procedure–Powder/ChipSamplesDeterminetheinstrument...
【详情】氧氮氢分析仪的常见故障及解决方法供电正常、通讯正常,点击确认键后分析仪不工作。这是没有水流,炉子温度...
【详情】