Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Add approximately 0.05 g of *** Graphite Powder to a***Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen,the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 3b through 3i a minimum of three times for each calibration/drift standard used.
***电极炉头设计,提高脉冲炉热交换效率。江苏美国分析仪哪家好Procedure - Solid Samples
Determine the instrument blank.
Repeat steps 2b through 2g a minimum of three times.
Set the blank following the procedure outlined in the operator's instruction manual.
Instrument calibration/drift correction.
Login a minimum of 3 Standard replicates.
Weigh approximately *** grams of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.
Note: LECO Reference Materia***o not require preparation.See preparation statement on the reference material certificate.
山西LECO分析仪价格氧氮氢分析仪选哪家?给你推荐上海禹重科技!
Sample Preparation
Sampling and sample preparation of refractory metals such as titanium and zirconium is somewhat different from that of steel. Unlike steel samples, hydrogen is not as mobile in this group of materials; therefore, storage in liquid nitrogen or dry ice is not required. However, it is important to keep the sample cool when cutting or sectioning. Sample preparation for oxygen and nitrogen determination has beendifferent from that of hydrogen determination.
Procedure – Powder/Chip Samples
Prepare the instrument as outlined in the operator's instruction manual.
Login a minimum of three Blank reps.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open. Insert a *** Nickel Capsule (leave capsule open) into the open port at top of loading head.
Note: samples using automation should be placed in the appropriate autoloader position before starting the analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.
Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
Place the Nickel Capsule containing the sample into the open port at the top of the loading head.
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode either manually or remove the crucible and press the analyze button to clean with an automatic cleaner if applicable.
氧氮氢分析仪怎么选?欢迎联系上海禹重科技,我司产品质量保证,欢迎咨询!江苏美国分析仪哪家好
Procedure - Solid Samples
Instrument calibration/drift correction.
Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 3b through 3j a minimum of three times for each calibration/drift standard used.
Calibrate/drift following the procedure outlined in the operator's instruction manual.
氧氮氢分析仪的常见故障及解决方法仪器启动时显示。没有水流。系统分析电流切断,分析停止。这是水流探测器...
【详情】736Series:Oxygen/NitrogenbyInertGasFusionThe736fam
【详情】美国LECO力可氧氮氢分析仪ONH836氧氮氢联测仪系列软件部分1、用户友好界面的Cornersto...
【详情】Procedure–Powder/ChipSamplesInstrumentcalibration/...
【详情】LECO—YoursourcefortotalanalyticalsolutionsCSCS844a...
【详情】InstrumentHighlightsandFeaturesFeaturesandBenefits
【详情】仪器的亮点和优势:特点和优势·炉头区域LED照明系统·炉头快速拆装的粉尘过滤装置·装有自动旁路阀的可...
【详情】Procedure–Powder/ChipSamplesDeterminetheinstrument...
【详情】氧氮氢分析仪的常见故障及解决方法供电正常、通讯正常,点击确认键后分析仪不工作。这是没有水流,炉子温度...
【详情】