Procedure–Powder/ChipSamplesDeterminetheinstrumentblank.PresstheAnalyzebuttonontheinstrumentscreen,thelowerelectrodewillcloseandtheanalysissequencewillstartandendautomatically.Repeatsteps2bthrough2gaminimumofthreetimes.Settheblankfollowingtheprocedureoutlinedintheoperatorsinstructionmanual.Instrumentcalibration/driftcorrection.LoginaminimumofthreeStandardreps.Weighapproximately***gramsofacalibration/driftstandardintoa***NickelCapsule;enterthemassandstandardidentificationintoappropriaterepfields.Ifapplicable,placethenickelcapsuleintheappropriateautoloaderposition.Note:LECOReferenceMateria***onotrequirepreparation.Seepreparationstatementonthereferencematerialcertificate.SolidStandardsmaybeusedtocalibratewhenchiporpowderstandardsarenotavailable.分析仪,就选上海禹重实业有限公司,欢迎客户来电!浙江专业元素分析仪咨询报价
氧氮氢分析仪的操作过程氧氮氢分析仪可根据的用户的要求定制。该分析仪配备两个不同灵敏度的探测器,在更多的应用中保证了准确度。操作简单而且安全。在样品称重以后,样品的重量数据通过电子界面传导至相连的电脑中,也可以在软件界面中手动输入样品的重量。样品被放置在水平放置的炉子中的低温区域。分析开始以后,炉子开始向上转动,然后样品逐渐落入加热区。通过添加载体气体氮气,氢气被扩散出来并且随载体被传导至一个高灵敏度的热传导测量池。通常一次测量的时间在3到15分钟之间。在分析的过程中设备可以显示测试信号和设备的参数,从信号的评估到结果在界面上的显示全部自动完成;而测量数据则被传导至LIMS(实验室信息管理系统)中。无需过多的保养和维护,而除尘器和净化管的维护工作也方便。广东进口分析仪上海禹重实业有限公司为您提供分析仪,有需求可以来电咨询!
氧氮氢分析仪的常见故障及解决方法仪器启动时显示。没有水流。系统分析电流切断,分析停止。这是水流探测器不正常,水泵不工作,管道堵塞。将仪器的右面板取下,观察水流探测器,用手挤压补水塑料水瓶,发现水流正常,显示正常,证明水流探测器正常,管道畅通。启动循环水泵,但分析仪显示没有水流,此时可判定水泵不正常。打开水泵转子密封口,启动泵发现电机正常运转,此时关闭进水,拆下水泵,发现叶轮脱落。经了解,判定是由于外部冷却水停水,仪器内循环水温过高(水温应≤70℃),致使叶轮(叶轮材料PVC)热胀并脱轴。用粘合剂粘合叶轮后再粘于叶轮轴上,待粘合剂凝固后试车,仪器运行正常。
Procedure–Powder/ChipSamplesInstrumentcalibration/driftcorrection.Calibrate/driftfollowingtheprocedureoutlinedintheoperatorsinstructionmanual.AnalyzeSamples.LoginSamplewiththeappropriatenumberofreps.Weighapproximately***gramsofappropriatelypreparedsampleintoa***NickelCapsule,entermassandsampleidentificationintoappropriaterepfields.Ifapplicable,placethenickelcapsuleintheappropriateautoloaderposition.Repeatsteps3cthrough3iforsampleanalysis.上海禹重实业有限公司为您提供分析仪,欢迎您的来电哦!
Oxygen,Nitrogen,andHydrogeninRefractoryMetalsSummaryTitaniumisametalthatcanbecombinedwithelementssuchasaluminum,vanadium,molybdenum,andtintoproducehigh-strength,lowdensity,andcorrosion-resistantalloys.Titaniumalloysareusedbythemilitary,medicaldevices,sportinggoods,andaerospaceindustriesbecauseoftheseproperties,andduetothestrictdemandsoftheseindustries,effortneedstobetakentoassurethatthematerialmeetsthehighestofqualitystandards.上海禹重实业有限公司致力于提供分析仪,有想法可以来我司咨询。浙江美国LECO分析仪优惠价格
上海禹重实业有限公司为您提供分析仪,有需要可以联系我司哦!浙江专业元素分析仪咨询报价
Procedure-SolidSamplesDeterminetheinstrumentblank.Repeatsteps2bthrough2gaminimumofthreetimes.Settheblankfollowingtheprocedureoutlinedintheoperatorsinstructionmanual.Instrumentcalibration/driftcorrection.Loginaminimumof3Standardreplicates.Weighapproximately***gramsofacalibration/driftstandard,enterthemassandstandardidentificationintoappropriatereplicatefields.Note:LECOReferenceMateria***onotrequirepreparation.Seepreparationstatementonthereferencematerialcertificate.浙江专业元素分析仪咨询报价
氧氮氢分析仪的常见故障及解决方法仪器启动时显示。没有水流。系统分析电流切断,分析停止。这是水流探测器...
【详情】736Series:Oxygen/NitrogenbyInertGasFusionThe736fam
【详情】美国LECO力可氧氮氢分析仪ONH836氧氮氢联测仪系列软件部分1、用户友好界面的Cornersto...
【详情】Procedure–Powder/ChipSamplesInstrumentcalibration/...
【详情】LECO—YoursourcefortotalanalyticalsolutionsCSCS844a...
【详情】InstrumentHighlightsandFeaturesFeaturesandBenefits
【详情】仪器的亮点和优势:特点和优势·炉头区域LED照明系统·炉头快速拆装的粉尘过滤装置·装有自动旁路阀的可...
【详情】Procedure–Powder/ChipSamplesDeterminetheinstrument...
【详情】氧氮氢分析仪的常见故障及解决方法供电正常、通讯正常,点击确认键后分析仪不工作。这是没有水流,炉子温度...